Surgeries & Interventions

Medtronic’s Defibrillation Innovation

Medtronic EV ICD

A new era of implantable cardioverter-defibrillators (ICDs) may be on the horizon. In this premarket clinical study, Medtronic’s unique extravascular ICD effectively terminated acute and chronic life-threatening arrhythmias, while simultaneously hitting safety endpoints.

In contrast to traditional ICD leads which are inserted through veins into the heart, this device’s lead is placed directly behind the sternum and connected to a device implanted below the patient’s left armpit. By placing the lead beneath the breastbone but outside of the heart, the team reduced risks associated with transvenous ICDs such as lung collapse, heart perforation, and heart valve damage. 

In the prospective single-group study of patients with a class I or IIa indication for an ICD, 316 participants underwent an attempted implant, and 299 were discharged with a working extravascular ICD system.

  • Physicians successfully terminated an induced ventricular arrhythmia in 298 of 302 patients (98.7%).
  • At six months, 92.6% of participants had no major system or procedure-related complications.
  • Out of 18 recorded episodes of cardiac arrest, 100% were converted back to normal heart rhythms after delivering shocks.
  • 23 patients experienced major complications, the most common being dislodgement of the lead wire.
  • 29 patients experienced inappropriate shocks, at an average of 10.6 months after implantation, with most linked to over-sensing of the heart’s electrical signals.

The Takeaway

Overall, researchers found that the EV ICD achieved a defibrillation success rate of 98.7%—the original goal was 88%—and met all safety endpoints. Medtronic has received a green light from the FDA to continue studying the device while the agency reviews its premarket application.

Get twice-weekly insights on the biggest stories shaping cardiology.

You might also like

Electrophysiology August 21, 2025

Conduction-System Pacing Succeeds at Treating AV Block August 21, 2025

Conduction-system pacing could be poised to reshape standard care for atrioventricular (AV) block patients after the randomized CSPACE trial demonstrated its significant benefits over traditional right ventricular septal pacing. Taking a closer look at CSP’s potential, the CSPACE trial enrolled 202 patients with AV block at two Australian hospitals, randomizing them to conduction-system pacing or […]

Cardiology Pharmaceuticals August 18, 2025

Rivaroxaban Monotherapy Could Be Better for CAD + AFib August 18, 2025

New analysis from the AFIRE trial suggests rivaroxaban monotherapy might be emerging as a preferred antithrombotic strategy for patients with AFib and stable coronary artery disease, especially in older age groups. The Post Hoc AFIRE Analysis examined 2,215 Japanese patients with AFib and stable CAD, stratifying them into four age groups (<70, 70-74, 75-79, and […]

Surgeries & Interventions August 14, 2025

TEER Could Save Lives in AFMR Patients August 14, 2025

Shedding light on a new treatment for an understudied condition, data published in EHJ suggests that transcatheter edge-to-edge repair (TEER) delivers significant survival advantages in patients with atrial functional mitral regurgitation (FMR). The new OCEAN-Mitral/REVEAL-AFMR Analysis compared 441 TEER-treated patients against 640 medically managed controls with moderate or severe atrial FMR, stating a solid case […]